ANALYSIS OF VARIATION OF TIN
FILTER ON NOISE VALUES IN CT SCAN MASTOID PROTOCOL USING CT SCAN SINGLE SOURCE:
PHANTOM STUDY
Merry Suzana1, Kusworo
Adi2, Darmini3
Politeknik Kesehatan Kemenkes Semarang, Jawa Tengah,
Indonesia
|
KEYWORDS |
ABSTRACT |
|
ct scan mastoid, tin filter, noise, snr, nps, indoqct. |
Since its introduction in 1972, CT Scan technology has developed rapidly
from year to year, especially in reducing radiation dose; various
technologies have been developed, including dual-energy or dual source
technology with a technique of only a quarter of the gantry rotation. The aim
of this research is to find out and analyze tin
filter variations on noise in the mastoid CT scan protocol using a single
source CT scan: phantom study. The method used in this research is Pre-Experimental with Posttest-Only Control Design. The noise value research
results showed a difference in noise and SNR values with a p-value
(p<0.05), and there was a reduction in radiation dose of 41.95% compared
to the standard protocol. NPS values are at a frequency of 0.28, except for
the Sn100 protocol. Subjective Analysis by two radiologists with a kappa
value of 0.75 indicated that there was a moderate level of agreement. The Tin Filter Sn100 protocol can reduce the
radiation dose, and the resulting image quality is good enough for diagnostic
purposes, so the Tin Filter Sn100 protocol can be used as a Standard Operational Procedure for
Mastoid CT Scan examination at Hermina Depok Hospital. This study has
implications that may contribute to refining mastoid CT scanning protocols,
leading to improved image quality, reduced radiation exposure, and improved
diagnostic capabilities in clinical practice. |
|
DOI: 10.58860/ijsh.v2i9.92 |
|
Corresponding Author: Merry Suzana
E-mail: merrysuzana@gmail.com
INTRODUCTION
Since its introduction in 1972, CT Scan technology
has developed rapidly from year to year, especially in reducing radiation dose;
various technologies have been developed, including dual-energy or dual source
technology with a technique of only a quarter of the gantry rotation. Can
produce images that can be processed so that the X-ray exposure time can be
reduced, and then the radiation dose released can be reduced by up to a quarter
of a percent (Ertel et al., 2009; M. Lell et al., 2009; Leschka et al., 2009; Tacelli
et al., 2010), and some have developed filter back projection reconstruction techniques or
what is known as iterative reconstruction to adapt low dose images that are
identical to rough or noisy images. (Kalmar et al.,
2014; Singh et al., nd) .
Software technology such as Automatic Tube Current Modulation has also been
developed; this software can reduce the radiation dose during CT Scan examination exposure. This technology automatically
adjusts the tube current (mA) and adjusts attenuation variations in the
patient's body when the X-ray tube rotates from various angles, known as
angular tube current modulation (Mannudeep
K. Kalra, Michael M. Maher, Thomas L. Toth, Bernhard Schmidt, Bryan L.
Westerman, Hugh T. Morgan, 2004; McKenney et al., 2014; Mcnitt-gray, 2011). All of these radiation dose reduction technologies are very
useful in scanning organs with high radiosensitivity,
including
the organs in
the head area.
The organs in the head area have a high level of density, so
in practice, the CT Scan examination carried out on this part of the head
requires quite a large dose; apart from that, there are also parts of the organ
that require High-Resolution
image quality to make a diagnosis—a for example, in the mastoid air cell organ (Dexian
Tan et al., 2018). Mastoid Air Cells are small spaces
filled with air located in the mastoid process of the temporal bone. These
cells are part of the middle ear and are important for regulating air pressure
in the ear (Magnuson, 2003). The Mastoid Air Cell organ is close to organs that
have radiosensitive properties, such as the eye lens (Sowby, 1981) and the thyroid
gland. It has been researched that X-ray exposure in radiodiagnostics can
increase the risk of microcarcinoma in the thyroid organ (Kovalchuk &
Kolb, 2017). Therefore, during CT scanning, This
mastoid
air cell scan requires great care in determining the radiation dose used in the
examination, and this radiation dose is very closely related to the quality of
the resulting image.
These image qualities include Spatial, Contrast, and Temporal
Resolution (Romans, 2011). Of these three image qualities, the most
influential when the CT scan image is underexposed is Contrast Resolution. This
Contrast
Resolution can be assessed with a phantom containing objects; distinguishing very small
objects is difficult to do subjectively. Therefore, noise is important in
detecting large images—low-contrast resolution. Noise, SNR, and NPS are the most
complete analyses in terms of low-contrast resolution (SEERAM,
2016).
In 2017, a patient pre-filter system was developed by Siemens
Healthiners on the Single Source CT Scan system,
marketed under Tin Filter (Braun et al., 2015). This Tin Filter is the development of a pre-patient system
on the previous system known as the bowtie filter; this bowtie filter is the
basic filtration in the CT Scan system, which is
useful in forming X-ray beams. The principle of the Tin Filter is the same as when we
look at the sun with and without glasses; of course, it will be easy to see the
sunlight with glasses clearly to eliminate the dazzling effect. Tin Filter or Sn filter, Sn comes from the
Latin word for tin, namely stanum. The Tin Filter is
an additional filter placed in front of the bowtie filter to filter the beam
from the X-ray tube before it hits the patient's body. As a result, only
high-energy photons reach the patient, while very low energy will not be able
to penetrate this tin filter will not reach the patient and will be filtered so
that the radiation dose can be suppressed (Choi et al., 2020;
Hamann et al., 2017; MM Lell et al., 2015; Rajendran et al., 2020), by producing high photon energy can reduce noise and contrast
in the image, although this filter method can reduce exposure to ionizing radiation
reaching the patient, noise in the image has an important role in terms of
image quality and diagnostic performance (Mozaffary et al., 2019). This tin filter has a selection variation
according to the energy level used between Sn100, Sn110, Sn120, and Sn140; Sn
indicates the use of tin filters. For example, Sn100, at 100 kV energy, uses
tin filters as energy filters (Schüle et al., 2023 ).
In recent years, there have been studies related to
lead filters on CT scans with multiple sources on Thorax CT scan examination
with Covid diagnosis with lead filters using Sn150 kV studied by researchers (Agostini et al., 2021) . In 2020
researchers with low-dose CT results with spectral and ADMIRE3 formation were
able to produce acceptable image quality for the evaluation of Covid19 patients
and significantly reduce dose and movement artifacts (Agostini et al., 2021) . The use of tin
filters on Sn100 kV and Sn140 kV at the shoulder joint by Yun Seok Choi et al.
2020 (Choi et al., 2020) with the results of Shoulder CT Scan with Sn140 kV can reduce
radiation dose by around 70 – 60% and image noise when maintaining contrast
images compared to conventional protocols, Tin filter with a combination of the
high pitch in detecting ureteral stones by Gu Mu Yang
Zhang et al. in 2017 (Zhang et al., 2017)
with the results of high
pitch C T Scan Pelvic Abdomen at Sn150kV can substantially reduce radiation
exposure when detecting stones in the urinary tract, and the application of
dose reduction and diagnostic performance of tin filters in colorectal cancer
patients by Koichiro Kimura et al. in 2022 (Kimura et al.,
2022) with the results of a 100
kV tin filter can reduce the radiation dose by around 89% compared to the
standard 120 kV protocol.
Hermina Depok Hospital has just installed the latest
CT Scan, and there is a pre-system filter in this CT Scan system. This is the first time it has been embedded in a
single-source CT Scan system, which previously only existed in dual-source
systems; this
new filter system can reduce the radiation dose by around 89% (Kimura et al., 2022), with a dose that is reduced quite drastically, of
course, it would be a shame if it is not used properly, however, if the
radiation dose is reduced so drastically, radiographers are still unsure. Using
the Tin Filter will increase image noise, which is caused by reducing the
number of photons reaching the photo object. Noise itself has an important role
in terms of image quality and to know for sure the results of the image
provided and how well it is in reducing the radiation dose with image noise
that is still within tolerance limits in diagnosing an abnormality in the
mastoid air cells. An inspection of Mastoid air cells is required to show the smallest structures of these
organs so that they can be seen well, especially the bones, malleus, and
incus.
Several radiologists at Hermina Hospital Depok complained about this
mastoid examination for small bone parts that are blurry and somewhat unclear in
that area;
this examination is carried out quite a lot for around 25% of the total
patients every month. Therefore, high-resolution image quality is required to establish a proper
diagnosis. The weakness of the standard bowtie filter pre-filter system is that
the output beam resulting from the collision is still mixed with super low
energy and other energies; therefore, organs with a high-density level in their
smallest structures will result in blooming artifacts.
In contrast to the tin filter principle, whose function is only to filter
low energy and pass only high energy used in scanning. This may be useful in
scanning organs with a high level of density to determine the quality of
the image produced and the level of noise produced by each variation of filter tin
given in the Mastoid CT
Scan protocol, considering that examinations using radiation must be as
efficient as possible and must conform to the ALARA principle.
Based on the background that has been written, it is
very interesting to carry out a more in-depth noise analysis regarding the
levels of noise values, Signal-to-Noise (SNR) and Texture Noise Power Spectrum
(NPS) and the resulting dose reduction for each tin filter variation (Sn100,
Sn110, Sn120 and Sn140) with the mastoid protocol on a water phantom object as
a baseline image, then scanned again with a rando phantom which has
specifications similar to a real human body. This is interesting to research; what
causes the noise produced at each energy still needs to be widely known, and
previous studies only examined one or two tin filter variations. Apart from
that, this tin filter system is something new in the single source CT scan
system, and of course, with a reduced radiation dose, it will produce low
noise, which is very contrary to the principle that if the radiation dose is
low or underexposed, you will get high noise and things like this can reduces
diagnostic performance. This research will be used as the basis for making the
Mastoid CT Scan Protocol SOP Low Dose. So, the aim of this research is is
to find out and analyze tin filter variations on
noise in the mastoid CT scan protocol using a single source CT scan: phantom
study.
METHOD
Pre-Experimental
Research with a Posttest-Only Control Design.
This research began with scanning Water Phantom in standard protocol and
followed by scanning variations of Tin Filter (Sn100,
Sn110, Sn120, and Sn140), then
assessed for noise and SNR. Noise texture or NPS was analyzed
using indoQCT Version 22
software. Then, after getting a baseline picture of which
protocol is good, the protocol is used as a scanning
protocol with phantom random objects, and
interobserver agreement is calculated.
This
research begins by preparing the Go Top and Water Phantom CT scan planes. Then,
adjust the position of the water phantom on the examination table guided by the
horizontal collimation lamp and the longitudinal collimation lamp. Position the
water phantom in the middle and set the gantry tilt to 0°. Data was collected
by scanning the water phantom without using a Tin Filter and Tin Filter
variations (Sn100, Sn120, Sn140). Set the exposure
parameters on the Siemens Somatom Go Top console
table. In accordance with the mastoid examination protocol, namely with kV
according to Sn Filter variations, mAs with Automatic Tube Control Modulation, tube rotation
0.6 seconds, detector collimation width 1.2 mm, image thickness 1.5 mm, B41s
kernel. Scanning is done from top to bottom, starting by making a program.
Researchers took scanning using an axial technique, taking several variations
of Sn Filter with each series taking ten slices from
the water phantom at slice numbers 12, 24, 36, 48, 60, 72, 84, 96, 108, and 120
and taking images of the water phantom without using tin. Filter by providing
an un-checklist in the enable/unenable and scanning
dialog boxes, and by using Tin Filter by selecting the Sn
Filter variation that will be used Sn100, Sn110, Sn120, and Sn140. Next, the
image resulting from the radiation is stored on the workstation. Next, the
image is analyzed and measured; noise and SNR values
by taking ROI (region of interest) are taken at points referring to Figure
1—Region of Interest (ROI) Measurement Points.

Figure 1. ROI
Measurement Point
In addition, noise evaluation was
also carried out using the second measurement NPS method, namely noise using
the Noise Power Spectrum method using IndoQCT software in terms of the size of
the noise and the texture of the noise itself, and the third Analysis, namely
the radiation dose in the form of CTDIvol with units of mGy recorded on the
workstation. After getting the analysis results and getting a good idea of the
Tin Filter protocol, the protocol is used for scanning phantom rando, which has
specifications similar to the original human body. Apart from that, noise
evaluation was also carried out using the NPS method, the second measurement
was noise using the Noise Power Spectrum method using IndoQCT software in terms
of the amount of noise and the texture of the noise itself, and the third
Analysis was radiation dose in the form of CTDIvol in mGy units recorded on the
workstation. After getting the analysis results and getting a good idea of the
Tin Filter protocol, the protocol is used for phantom rando scanning, which has
specifications similar to the original human body in Figure 2.

Figure 2. Phantom Rando
Then, the image quality from the
results of the phantom rando scanning is analyzed subjectively by a radiologist
at the hospital who will apply the Mastoid Low Dose Protocol SOP.
RESULTS
AND DISCUSSION
True -Experimental
research has been carried out with the Posttest-Only
Control Design. This research was carried out by scanning two types of
phantoms, namely water phantom and rando phantom, to
look for good image quality, which will be used as a new protocol and SOP for
Low Dose Mastoid CT scan examinations. This research was conducted at Hermina Hospital, Depok, with data
collection in April 2023.
This study
used images taken from water phantom scanning; the scanning
was carried out five times with
standard protocols/without filter tin, Sn100, Sn110, Sn120, and Sn140. From each of these protocols, 124 images were
obtained, and then ten images were selected as samples
representing each area level in the water phantom, namely the edge area. Next, ten images from each
standard protocol, Sn100, Sn110, Sn120, and Sn140, had their noise values
measured at 12 measurement points, as in Figure 4.2. These 12 points were
taken because, at each point, the area in the ROI
produced a different value. Therefore, 12 different points were taken in the hope that they could represent each area in the center and other
edges, and in this selected image, there were no artifacts to make it easier in the analysis process, namely in slices
12, 24, 36, 48, 60, 72, 84, 96, 108 and 120. The following are the results of
scanning water phantoms obtained on the standard protocol (a), Sn100 protocol
(b), Sn110 protocol (c), Sn120 protocol (d), and the Sn140 (e) protocol, as shown in Figure 3.
|
|
|
|
||
|
a |
b |
c |
||
|
|
|
|
|||
|
d |
e |
|
|||
Figure
3. Water Phantom Scanning Results

Figure
3.
Twelve Noise and NSR Value Measurement Points
After
being measured using ROI ( region of interest ), the measured noise value is
obtained from the standard deviation of the ROI measurement. Then, the 12
measurement points are averaged. The noise values of the ten images in each
image from each protocol are averaged with the following results:
Table 1 .
Noise Value
of Each Tin Filter Protocol
|
Protocol |
N |
Noise Average |
Standard Deviation |
Min Value |
Maximum Value |
|
Standard |
10 |
14.0600 |
0.40729 |
13,43 |
14.71 |
|
sn100 |
10 |
14.5990 |
0.34323 |
14,11 |
15,13 |
|
Sn110 |
10 |
13.5380 |
0.22744 |
13.23 |
14.03 |
|
sn120 |
10 |
12.7190 |
0.20760 |
12.34 |
13.05 |
|
Sn140 |
10 |
12.6670 |
0 .29010 |
12.21 |
13.03 |
From Table 1, it can be seen that in the standard protocol, the average
noise value is 14.0600. From each tin
filter protocol, the lowest average noise value is in the Sn140
protocol of 12.6670, and the highest average noise value is in the Sn100
protocol of 14.5990.

Figure
5. Noise
From Figure 5, it can be seen that in
the standard protocol, the noise value is 14.059. In the tin filter variation
protocol, the highest noise value produced was in the Sn100 protocol at 14.598,
and the lowest was in the Sn 140 protocol at 12.667.
The use of variations in the filter shows that as the energy used
increases,
the noise value decreases from Sn100 to Sn140. Then, the ANOVA test was continued on each Tin
Filter protocol with a sig value of 0.000<0.05, so it can be concluded that
the average noise value of these five Tin Filter protocols is significantly different.
Then, proceed with post hoc ANOVA to analyze each variance between groups.
In this experimental research, only the average noise values of the Sn140 and
Sn120 protocols are the same, while the average noise values in the other
protocols are different; thus, the Tin Filter variable The noise value only has
a significant effect on the Standard, Sn100, and Sn110 protocols.
The assumption in image assessment is that the
higher the noise value, the rougher the resulting image (Bushberg et al., 2012; Romans, 2011; SEERAM, 2016;
Tack et al., 2012). This noise
value is influenced by three main factors: the number of X-ray photons detected,
the physical limitations of the equipment, and reconstruction parameters. In
this study, the reconstruction parameters and devices used were also the same
between each protocol, but there were differences in using the Tin Filter. The
number or quality of X-ray photons the detector captures differs for each
protocol. Suppose you look at the average noise value in Table 4.1. In that
case, it can be seen that between the standard mastoid CT Scan protocol
(14,060) and the Tin Filter Sn100 protocol (14,599), the values are not too
different; however, in the description of the Tin Filter Sn100 protocol, it is
a bit rougher. Suppose you look at the comparison between each Tin Filter
protocol Sn100 (14,599), Sn110 (13,538), Sn120 (12,719), and Sn140 (12,667). In
that case, the noise value decreases as the amount of energy (kV) used
increases; this proves that the higher the energy ( kV) used when using a Tin
Filter, the lower the noise produced; this is confirmed by previous research by
Yun Seok Choi in 2020 on CT Shoulder Arthrography with Tin Filter at energies
of 100 kV and 140 kV with lower noise results than conventional protocols, and
at an energy of 100 kV or Sn100 the radiation dose is more efficient (Choi et al., 2020), this proves that low radiation doses do not always
produce high image noise, this is also the same as research by Sonja Gordic et
al. in 2014 on Ultralow-Dose regarding Chest Computed Tomography in Pulmonary
Nodule detection that the use of this Tin Filter can reduce the noise produced.
By using a Tin Filter, the energy that passes through the Tin Filter is only
high energy, so in this Tin Filter protocol, the resulting image resolution is
better than the standard protocol. Noise level indicates the radiation dose
received; when using a bowtie filter or standard protocol/sign tin filter,
image noise tends to be greater in the peripheral area, indicating a high local
patient dose in that area (Woods & Brehm, nd). Use of Tin Filter with noise closely related to the
X-ray photons used. Photon scattering is another important interaction effect
in CT imaging besides photon absorption via the photoelectric effect. Photon
scattering creates noise in the primary signal, and a high primary scattering
ratio results in image artifacts (Woods & Brehm, nd). This is the case when many scattered photons from
a short cutoff length contribute to the signal for a small number of primary
photons from a long cutoff length. To increase the scatter-to-primary ratio,
the number of incident photons for peripheral body parts is reduced, and thus,
the number of scattered photons (Woods & Brehm, nd). Thus, this Tin Filter makes the noise between the
edge and the center more homogeneous.
Next, the SNR value (signal-to-noise ratio) obtained
from the HU value measured by the ROI is divided by the standard deviation of
the measured ROI. This SNR value was also measured at the same 12 points as the
noise, and then the SNR values from 12 points from 10 images in each protocol
were averaged with the following values:
Table 1. SNR value of each Tin
Filter Protocol r
|
Protocol |
N |
Average SNR |
Standard Deviation |
Mark Minimum |
Maximum Value |
|
Standard |
10 |
0.4530 |
0.02541 |
0.40 |
0.48 |
|
sn100 |
10 |
0.4610 |
0.03035 |
0.41 |
0.51 |
|
Sn110 |
10 |
0.4710 |
0.01595 |
0.44 |
0.49 |
|
sn120 |
10 |
0.5280 |
0.03425 |
0.46 |
0.58 |
|
sn140 |
10 |
0.5150 |
0.01354 |
0.49 |
0.54 |
Table 3 shows the average SNR value on the standard
protocol is 0.4530. The highest average SNR value on the tin filter variation
protocol is on the Sn120 protocol, with an average noise value of 0.5280. The
lowest average SNR value is on the Sn100 protocol of 0.4610.

Figure 4. SNR Value Graph
Figure 6 shows that the SNR value for tin filter
variations increases, with the highest SNR value being on the SN120 at 0.529
and the lowest SNR value being on the Sn100 protocol at 0.460. The SNR value
for the SN100 protocol is 0.460, almost similar to the SNR value for the
standard protocol, which is 0.452. From the use of tin filter variations, the SNR value also increases from
Sn100 to Sn140 as the energy used
increases. Furthermore, the ANOVA test
on each Tin Filter protocol resulted in a sig value of 0.000<0.05, so it can
be concluded that the average SNR value of these five Tin Filter protocols is
significantly different. Then,
proceed with post hoc ANOVA to analyze each variance between groups. In this
experimental research, only the average SNR value of the Standard, Sn100, and
Sn110 protocols is the same, while the average SNR value of the Sn140 and Sn120
protocols is the same; thus, variable Tin Filter SNR value has two protocol
groups that produce the same SNR value.
Then, proceed with post hoc ANOVA to analyze each
variance between groups. In this experimental research, only the average SNR
value of the Standard, Sn100, and Sn110 protocols is the same, while the
average SNR value of the Sn140 and Sn120 protocols is the same; thus, variable
Tin Filter SNR value has two protocol groups that produce the same SNR value.
The SNR value better reflects the signal from each image
pixel, the average background signal value, and the deviation from the uniform
background in the image. The higher the SNR value, the better the image quality
(Bushberg
et al., 2012; El-Khoury et al., 2004; Romans, 2011; SEERAM, 2016; Tack et al.,
2012). The SNR value in this study is in Table 4.2 in the standard protocol
(0.453), and the t in protocol f filter Sn100 (0.461). The value is not too different, but
the SNR value of the Tin Filter Sn100 protocol is higher than the standard
protocol. If we look at each Tin Filter protocol Sn100 (0.461), Sn110 (0.471),
Sn120 (0.528), and Sn140 (0.515), the SNR value is higher; this proves that the
higher the energy (kV) used will increase the value. The resulting SNR. This is
supported by previous research by Andrea Agostini et al. in 2020 regarding
third-generation interactive reconstruction with high pitch, low dose Thorax CT
Scan protocol with Tin Filter for beamforming at 100 kV with the result that
the SNR value increased in the protocol using Tin Filter (Agostini
et al., 2021), and in research by Patricia
Dewes et al., in 2016 regarding Low-dose Abdominal CT Scan for the detection of
stones in the urinary tract - the effect of adding spectral shaping results in
an increase in the SNR value in the protocol that uses the Tin Filter, it will,
but at Sn150 energy the values are not significantly different. This proves
that at energies of 140 kV and above, there will be a decrease in the SNR
value, the same as in this research on Sn140; the SNR value also decreased. The
SNR value in Holger Haubenreisser's research, 2015 regarding the CT Scan of the
Thorax on the third generation dual-source CT Scan using a tin filter for
spectral shaping at 100 kVp, stated that the SNR value increased in parts of
the Aorta, Adipose tissue Lung, Trachea, but in the Supraspinatus muscle does
not increase. This increase in the SNR value is important to assess because SNR
is one of the most meaningful metrics that describes the conspicuity of an
object or how well it will appear to an observer (Bushberg
et al., 2012).
After measuring the noise and SNR values, the radiation dose recorded from
the results of scanning the water phantom in the form of CTDIvol and DLP is
recorded as follows:
Table 2. Radiation Dose Results
from Each Tin Filter Protocol
|
CTDIvol |
DLP |
||
|
Standard |
16.71 |
110 |
|
|
Sn100 |
9.70 |
81 |
|
|
Sn110 |
13.42 |
113 |
|
|
sn120 |
16.53 |
139 |
|
|
sn140 |
20,24 |
170 |
In Table 5,
each CTDIvol radiation dose produced using the tin
filter variation protocol tends to increase from Sn100 to Sn140, with the
lowest CTDIvol result being in the Sn100 protocol at
9.70 and the highest CDTIvol result being in the
SN140 protocol at 20. .24 . From the
results of the recorded radiation dose, it can be seen that the standard/no tin
filter protocol tends to be greater, namely 16.71, compared to the Sn100
protocol of 9.70.
After analyzing the recorded noise, SNR, and radiation dose
values, proceed with analyzing the NPS (noise power spectrum) with indoQCT
software version 22 by uploading an image that represents the results of the
water phantom scanning, namely on slice number 66, which is at the middle level
of the water phantom in each standard protocol, Sn100, Sn110, Sn120, and Sn140.
This NPS value is obtained from the ROI in a homogeneous area shown in Figure
7,

Figure 5. ROI area in IndoQCT software
Then,
the ROI of each protocol is stored in the form of a 300x300 pixel cropped
square shown in Figure 8. This is done to assess the characteristics of the
noise by assessing the resulting texture noise in more detail.
|
|
|
|
|
a |
B |
c |
|
|
|
|
d |
e |
Figure 6. Results: Each
protocol's sample image is cut to 300 x 300 pixels. a).
Standard, b). Sn100, c). Sn110, d).
Sn120 and e). sn140
Next, NPS measurements were carried
out by carrying out a Fourier transformation on the average pixel value in each
ROI. Fourier transformation is a model that moves the spatial or time domains
into the frequency domain. Fourier transformation is a process widely used to
move the domain of a function or object into the frequency domain, and then
after carrying out the Fourier transformation, it produces a 2-dimensional
image, which can be seen in Figure 9. The 2-dimensional image in the image shows the
texture of the focus point, which varies in each image.
|
|
|
|
|
a |
b |
c |
|
|
|
|
d |
e |
Figure 7. 2D image results from the Fourier transform of the average pixel in
each
ROI . a). Standard, b). Sn100, c). Sn110, d). Sn120 and e). Sn140
After uploading an image of
the water phantom results and putting the ROI on the image, click calculate.
Then, the results can be saved, as shown in Figure 10. Then, the
NPS calculation results from the IndoQCT software can
be exported in Excel format. The following is a recap of the NPS results in Table 6 :

Figure 8.
IndoQCT displays NPS calculations
Table 3. NPS value of each Tin Filter Protocol
|
Protocol |
Frequency |
Peak NPS Value |
|
Standard |
0.28 |
430.65 |
|
Sn100 |
0.2 |
391.80 |
|
Sn110 |
0.28 |
255.52 |
|
Sn120 |
0.28 |
330.89 |
|
Sn140 |
0.28 |
253.74 |
From Table 6, it can be seen that among all
the protocols, the frequency is 0.28, but the Sn100 protocol is at a frequency
of 0.2, and this difference is not too big. Of each protocol used, the standard
protocol has the highest peak value and decreases as the energy used increases.
The highest peak value was 430.65 in the standard protocol/without tin filter,
and the lowest was in Sn140, with a value of 253.74. The NPS value data from
the IndoQCT software is exported in Excel format, and then a
graph is created as below:

Figure
9. Graph of NPS Values for Each Tin
Filter Protocol
If simplified by just selecting the peak NPS value, it would be as
follows:

Figure 10.
Peak NPS Value of Each Tin Filter Protocol
The noise power spectrum can generally describe the
noise properties of an imaging system. This NPS is a more comprehensive
description than the standard deviation of a pixel or what we know as image
noise value. This NPS explains the characteristics of the magnitude and spatial
frequency of noise in CT scan images, which plays an important role in
analyzing and optimizing the imaging system's performance. This NPS is most
often associated with other parameters to assess image quality, and this NPS
has been commonly used in the development, characterization, optimization, and
comparison of various new imaging technologies. The tin filter protocol analyzes the NPS value
from the scanning water phantom using IndoQCT version 22 software. This software
is very simple in application and easy to understand for new users. This
software has been used in various research, such as research on automatic
validation of gantry tilt (Noviliawati et al., 2021), research on
automation of CT Number linearity measurements on ACR phantoms (Anam et al., 2023), research on
evaluating automatic measurement of slice profile sensitivity (Widyanti et al., 2023), and research
on the comparison of NPS and MTF in iterative reconstruction and deep learning
reconstruction (Setiawan et al.,
2023). The first step in the IndoQCT software is to upload images from each research
protocol. From the initial image (spatial domain), it is converted into a
frequency domain in the form of a 2D image that has been Fourier transformed as
in Figure 4.5; the difference in the 2D image at a glance is not that big. The
difference is visible. The 2D image results in each protocol are still in good
condition because the 2D noise spectrum image has a tight focus point. Figure
13 below shows a good reference noise spectrum image. The further to the left
the image the noise spectrum is, the lower the CT scan equipment used in this
research is still in good condition.
The peak value is in the standard protocol, but each
of these protocols is at almost the same frequency, except for the Tin Filter
Sn100 protocol, which is at a frequency of 0.2. The relationship between
frequency and NPS value is that the higher the frequency, the noise produced,
the lower it will be, or in other words, the better the texture noise produced (Bushberg et al., 2012; SEERAM, 2016). If we look at graphs 4.9 and 4.10, the graphs are
almost similar in shape to the standard protocol / without a tin filter, namely the Tin
Filter Sn100 protocol.

Figure 11. Noise Level Reference
After analyzing the value of noise, SNR, and NPS on
the water phantom, the best protocol was chosen with a noise level that was
almost similar to the standard protocol (14.0600 vs. 14.5990) but with a
much-reduced dose than the standard protocol (16.71 vs. 9.70), namely the tin
filter protocol Sn100 and the SNR value is close to the standard value (0.4530
vs. 0.4610) and the texture noise or NPS is at an almost similar frequency
(0.28 vs. 0.20).
Noise, SNR, and NPS analysis on
this water phantom image found that the Tin Filter Sn100 protocol is very good
compared to the others, taking into account the noise value, which has the
highest value (14,599) compared to the Tin Filter Sn110 protocol (13,538),
Sn120 (12,719 ) and Sn140 (12,667). The Sn100 protocol is most similar to the
standard protocol (14.06), but a low or high noise value is not a guarantee
that the CT scan image produced is good or not, but the noise value is a
reference for the image to be able to diagnose a pathological disorder or not.
In the post hoc ANOVA analysis, the noise value proves that the noise value in
the Sn140 and Sn120 protocols is the same, so the effect of the tin filter on
the noise value only affects Sn100 and Sn110. In the Analysis of the SNR value,
this SNR reflects an original picture of the object. The higher the SNR value,
the better it reflects the original object; the SNR value in the protocol Sn100
(0.463), Sn110 (0.471), Sn120 (0.52), and The Sn140 (0.515) increases as the
number of energy increases, but when using Sn140 the SNR value drops and in the
post hoc anova analysis there are two groups of protocols that produce the same
SNR, namely Standard, Sn100 and Sn110 vs. Sn120 and Sn140 and at Analysis of
the Sn100 NPS value has a slight difference with the standard protocol in terms
of frequency, in the standard protocol it is at a frequency of 0.28 and in the
Tin Filter Sn100 protocol it is at a frequency of 0.2. However, the noise
produced is still better than the standard protocol, and the most important
aspect is the radiation dose; from the standard protocol to the Tin Filter
Sn100 protocol, the radiation dose is reduced by 41.95%. This is similar to
previous research by K. Kimura et al. in 2022 regarding the radiation dose and
diagnostic performance of using Tin filters in colorectal cancer patients on
the Tin Filter Sn100 protocol, which can substantially reduce the radiation
dose by 89%, but the area scanned for abdominal organs so that these different
density levels will affect the penetrating power of X-ray photons that reach
the detector. Therefore, the Sn100 protocol was chosen and tested again on a
rando phantom object; this rando phantom has specifications similar to real
humans, and its attenuation level is equivalent to tissue for X-ray diagnostics
(Shrimpton et al., 1981). With the above considerations, the Sn100 protocol
was tested with a different scanning object, namely the rando phantom object
whose specifications are very similar to real humans; this phantom has been
used in several studies, including research by Ulla Nikupaavo et al. in 2015,
regarding Lens Dose in Routine Head CT: comparing the differences between
optimization methods and Anthropomorphic Phantoms or phantom rando, with the
result that gantry tilting can reduce the radiation dose to the eye lens (Serhal et al., 2001) and research on radiation dose by Charbel Bou Serhal
et al. in 2001 by utilizing this type of phantom as a scanning object (Serhal et al. ., 2001) The results of the Rando phantom scanning are shown in three image slices which represent
the results of the Sn100 protocol scanning at the base, mid and apex levels of
the temporal bone as shown in Figure 14.



Figure
12. Mastoid Air Cell Scanning
Results in the Protocol
Tin Filter Sn100 with
Phantom Rando Object
Then, the image results were assessed subjectively by asking seven questions by
two radiologists with more than ten years of experience. About image
quality, noise, sharpness, anatomical boundaries, anatomical parts of the
cochlear apical turn, malleus incus, and cochlear nerve opening. This subjective assessment
is carried out using a Totoku brand medical grade monitor screen with a reference to the
image to be assessed as in Figure 15.




Figure 15. 1Reference to the mastoid image in the questionnaire
However, complete image data is also provided for processing in MPR mode
to maximize assessment. The following is a recap of the subjective assessment
results by two respondents, which are shown in the table below:
Table 6. Recap of Subjective Assessment Questionnaire Data from 7
Questions
|
Respondents |
Question |
Total |
||||||
|
1 |
2 |
3 |
4 |
5 |
6 |
7 |
||
|
Radiologist 1 |
4 |
5 |
4 |
5 |
3 |
4 |
5 |
30 |
|
Radiologist 2 |
4 |
5 |
4 |
5 |
4 |
4 |
5 |
31 |
In table 6 the
first question about image quality from the two radiologists gave a score of 4
meaning that the two radiologists agreed that there were few artifacts but were able to show anatomical structures in
detail, in the second question about the noise produced the two radiologists
gave a score of 5 meaning the picture produced is Excellent Image without artifacts and clearly shows the anatomical structure in
detail, on the third question about the sharpness of the image the two
radiologists gave a score of 4 meaning that there were few artifacts
but were able to show the anatomical structure in detail, on the fourth
question about the anatomical boundaries the two radiologists gave a score of 5
meaning Excellent Image without artifacts and clearly
shows the anatomical structure in detail, on the fifth question about the coclear apical turn anatomy the two radiologists gave a
different score i.e. 3 means that there are artifacts but not
significant but can still show anatomical structures well and can still be diagnosed
, while radiologist 2 gives a score of 4 meaning that there are few artifacts but are able to show anatomical structures in
detail , on the sixth question about the anatomy of the
malleus incuss the two radiologists give score 4 and
on the seventh cochlear nerve opening question both radiologists gave a score
of 5 meaning Excellent
image without artefacts and clearly shows the anatomical structure in detail .
Overall, the subjective assessment of the two radiologists agreed that the mastoid CT scan protocol
with Sn100 could provide better and more informative image quality results so
that it could assist radiologists in making a diagnosis. The resulting
kappa value is 0.75 with a significant value of 0.013, indicating that the
coefficient value indicates a correlation or level of agreement at a moderate
level (McKenzie & Mahnken, 2023). subjectively, namely noise, sharpness, anatomical
boundaries, and the smallest anatomical parts, namely the cochlear apical turn,
malleus incus, and cochlear nerve opening. The anatomical parts assessed by the
observer are very important in supporting guiding surgery in the installation
of cochlear implants, so high-resolution images are needed here (Rodrigues et al., 2020). This mastoid CT scan image was also used in research by
Henrique Rodrigues, 2020 regarding Mastoid, middle ear, and inner ear analysis
on CT scan – possible contribution to cadaver identification, with CT scan
imaging of the Mastoid in visual assessment of the mastoid examination proving
to be a powerful and reliable approach to identify unique bone features and
contribute to human identification. In research by Khalid Hindi et al. in 2014
regarding Mastoid pneumatization of mastoid air cells and other parts of the
temporal bone, it was recorded as symmetrical in more than 75%. Pre-operative
assessment of the temporal bone and PNS with a CT scan can help evaluate
anatomical landmarks and reduce the possibility of surgical complications
because the image can clearly show the 3D structure (Hindi et al., 2014). Level of resolution on CT Scan This Mastoid needs
to be improved. Therefore, using a Tin Filter is a must to improve image
quality. In the future, imaging technologies should be developed to
non-invasively assess the structure and physiology of the middle ear/eustachian
tube in relation to their role in the pathogenesis of otitis media (Alper et al., 2017).
This Tin Filter was widely used in various
examinations, such as in previous research by Katharina Martini in 2015 (Martini et al., 2015) on the thorax protocol; it was proven that the tin
filter combined with iterative reconstruction was effective in detecting solid
and non-solid nodules with a specificity of 85 .7% and in research by Andrew D.
McQuiston et al., in 2016 the ability to detect the amount of calcium in
coronary blood vessels using the Sn100 protocol was proven to be accurate in
quantifying calcium scoring without correction for the HU threshold and in
research by Yun Seok Choi et al., in 2020 (Choi et al., 2020) in an arthrography examination proved that the use
of the Sn100 and Sn140 tin filter protocols was proven to reduce radiation dose
by 70% and 60% compared to the usual conventional protocol without reducing
image quality.
The Tin Filter protocol is very useful in mastoid CT
scan examinations because it is proven to reduce the radiation dose from the
standard protocol/without tin filter to the Sn100 protocol by 41.95%, and the
resulting image quality is quite good with noise values that are not much
different and sufficient to diagnostic needs and again the noise generated by
this protocol is still within reasonable limits because the increase in noise
produced by this Tin Filter variation is not too high compared to the standard
protocol/without tin filter, so this Tin Filter protocol can be used as a
permanent protocol and set as an inspection SOP Mastoid CT Scan at Hermina
Hospital Depok. The limitation of this research is that it cannot be carried
out directly on patients because it must consider the radiation dose given
during the examination, and in testing with a rando phantom, the pneumatization
image of the mastoid organ is not visible well because the preservation process
in the phantom so that the cavities in the Mastoid are already filled. Fluid.
CONCLUSION
Based on the results of research to analyze the
level of noise and reduction in radiation dose produced in each variation of
the tin filter in the mastoid CT scan protocol which will be used as a
reference in making SOPs, it can be concluded that there are differences in the
level of noise produced in each variation of the Tin Filter (Sn100, Sn110,
Sn120 and Sn140) given in the mastoid protocol with a p-value ( p <0.05) of
76.330 and there was a reduction in radiation dose from the standard protocol
to the Tin filter Sn100 protocol of 41.95%. The noise value decreases as the
amount of energy (kV) used increases; this proves that the higher the energy
(kV) used when using a Tin Filter, the lower the noise produced. The higher the
energy (kV) used in the Tin Filter protocol, the higher the SNR value produced;
however, at Sn140, the SNR value decreases so that at energies of 120 kV and
above, this Tin Filter has a certain tolerance limit. The highest NPS peak
value is in the standard protocol/without tin filter and decreases with the
energy used, but each of these protocols is at almost the same frequency,
except for the Tin Filter Sn100 protocol, which is at a frequency of 0.2. This
means that the resulting texture noise is similar. The radiation dose resulting
from the Standard Protocol/without tin filter, the CTDIvol radiation dose was
16.71 mGy with a DLP value of 110 mGy.cm; in the Tin Filter Sn100 Protocol the
CTDIvol radiation dose was 9.70 mGy with a DLP value of 81 mGy.cm, in the
Protocol The Tin Filter Sn110 radiation dose is 13.42 CTDIvol with a DLP value
of 113 mGy.cm, in the Tin Filter Sn120 Protocol the radiation dose is 16.53
CTDIvol with a DLP value of 139 mGy.cm and in the Tin Filter Sn140 Protocol the
radiation dose is 20.24 CTDIvol mGy with a DLP value of 170 mGy.cm. The higher
the energy used, the lower the radiation dose produced using a Tin Filter. The
image quality of the phantom rando object is of good quality and can show small
anatomical structures such as the anatomical parts of the cochlear apical turn,
malleus incus, and cochlear nerve opening. The results of the kohen kappa
between the two respondents show a kappa value of 0.75 with a significant value
of 0.013, indicating that the coefficient value indicates a correlation or
level of agreement at a moderate level. Overall, the two radiologists agreed
that the CT Scan Mastoid examination protocol with Sn100 can provide good image
quality results and assist in making a diagnosis.
REFERENCES
Agostini, A., Borgheresi, A., Carotti, M., Ottaviani, L.,
Badaloni, M., Floridi, C., & Giovagnoni, A. (2021). Third-generation
iterative reconstruction on a dual-source, high-pitch, low-dose chest CT
protocol with tin filter for spectral shaping at 100 kV: a study on a
small series of COVID-19 patients. Radiologia
Medica, 126(3), 388–398.
https://doi.org/10.1007/s11547-020-01298-5
Alper, C. M., Luntz, M.,
Takahashi, H., Ghadiali, S. N., Swarts, J. D., Teixeira, M. S., Csákányi, Z.,
Yehudai, N., Kania, R., & Poe, D. S. (2017). Panel 2:
Anatomy (Eustachian Tube, Middle Ear, and Mastoid—Anatomy, Physiology,
Pathophysiology, and Pathogenesis). Otolaryngology - Head and Neck Surgery
(United States), 156(4_suppl), S22–S40.
https://doi.org/10.1177/0194599816647959
Anam, C., Amilia, R., Naufal, A., Budi, W. S., Maya,
A. T., & Dougherty, G. (2023). The automated measurement of CT number
linearity using an ACR accreditation phantom. Biomedical Physics and
Engineering Express, 9(1), 17002.
https://doi.org/10.1088/2057-1976/aca9d5
Braun, F. M., Johnson, T. R. C., Sommer, W. H.,
Thierfelder, K. M., & Meinel, F. G. (2015). Chest CT using spectral
filtration: radiation dose, image quality, and spectrum of clinical utility. European
Radiology, 25(6), 1598–1606.
https://doi.org/10.1007/s00330-014-3559-1
Bushberg, J. T., Seibert, J. A., Leidholdt, E. M.,
Boone, J. M., & Goldschmidt, E. J. (2012). The Essential Physics of Medical
Imaging. In Medical Physics (Vol. 30, Issue 7).
https://doi.org/10.1118/1.1585033
Choi, Y. S., Choo, H. J., Lee, S. J., Kim, D. W., Han,
J. Y., & Kim, D. S. (2020). Computed tomography arthrography of the shoulder
with tin filter-based spectral shaping at 100 kV and 140 kV. Acta
Radiologica. https://doi.org/10.1177/0284185120965551
Dexian Tan, A., Ng, J. H., Lim, S. A., Low, D. Y. M.,
& Yuen, H. W. (2018). Classification of Temporal Bone Pneumatization on
High-Resolution Computed Tomography: Prevalence Patterns and Implications. Otolaryngology
- Head and Neck Surgery (United States), 159(4), 743–749.
https://doi.org/10.1177/0194599818778268
El-Khoury, G. Y., Bennett, D. L., & Ondr, G. J.
(2004). Multidetector-row computed tomography. In Journal of the American
Academy of Orthopaedic Surgeons (Vol. 12, Issue 1).
https://doi.org/10.1007/b139071
Ertel, D., Lell, M. M., Harig, F., Flohr, T., Schmidt,
B., & Kalender, W. A. (2009). Cardiac spiral dual-source CT with high
pitch: A feasibility study. European Radiology, 19(10),
2357–2362. https://doi.org/10.1007/s00330-009-1503-6
Hamann, A., Di, H. C., & Crm, C. T. (2017). Tin
Filter spectral shaping – a demonstration of Agatston equivalence.
Hindi, K., Alazzawi, S., Raman, R., Prepageran, N.,
& Rahmat, K. (2014). Pneumatization of Mastoid Air Cells, Temporal Bone,
Ethmoid and Sphenoid Sinuses. Any Correlation? Indian Journal of
Otolaryngology and Head and Neck Surgery, 66(4), 429–436.
https://doi.org/10.1007/s12070-014-0745-z
Kalmar, P. I., Quehenberger, F., Steiner, J., Lutfi,
A., Bohlsen, D., Talakic, E., Hassler, E. M., & Schöllnast, H. (2014). The
impact of iterative reconstruction on image quality and radiation dose in
thoracic and abdominal CT. European Journal of Radiology, 83(8),
1416–1420. https://doi.org/10.1016/j.ejrad.2014.05.017
Kimura, K., Fujioka, T., Mori, M., Adachi, T.,
Hiraishi, T., Hada, H., Ishikawa, T., & Tateishi, U. (2022). Dose Reduction
and Diagnostic Performance of Tin Filter–Based Spectral Shaping CT in Patients
with Colorectal Cancer. Tomography, 8(2), 1079–1089.
https://doi.org/10.3390/tomography8020088
Kovalchuk, A., & Kolb, B. (2017). Low dose radiation
effects on the brain–from mechanisms and behavioral outcomes to mitigation
strategies. Cell Cycle, 16(13), 1266–1270.
https://doi.org/10.1080/15384101.2017.1320003
Lell, M. M., May, M. S., Brand, M., Eller, A., Buder,
T., Hofmann, E., Uder, M., & Wuest, W. (2015). Imaging the parasinus region
with a third-generation dual-source CT and the effect of tin filtration on
image quality and radiation dose. American Journal of Neuroradiology, 36(7),
1225–1230. https://doi.org/10.3174/ajnr.A4270
Lell, M., Marwan, M., Schepis, T., Pflederer, T.,
Anders, K., Flohr, T., Allmendinger, T., Kalender, W., Ertel, D., Thierfelder,
C., Kuettner, A., Ropers, D., Daniel, W. G., & Achenbach, S. (2009).
Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT
angiography using dual source CT: Technique and initial experience. European
Radiology, 19(11), 2576–2583.
https://doi.org/10.1007/s00330-009-1558-4
Leschka, S., Stolzmann, P., Desbiolles, L.,
Baumueller, S., Goetti, R., Schertler, T., Scheffel, H., Plass, A., Falk, V.,
Feuchtner, G., Marincek, B., & Alkadhi, H. (2009). Diagnostic accuracy of
high-pitch dual-source CT for the assessment of coronary stenoses: First
experience. European Radiology, 19(12), 2896–2903.
https://doi.org/10.1007/s00330-009-1618-9
Magnuson, B. (2003). Functions of the mastoid cell
system: Auto-regulation of temperature and gas pressure. Journal of
Laryngology and Otology, 117(2), 99–103.
https://doi.org/10.1258/002221503762624512
Mannudeep K. Kalra, Michael M. Maher, Thomas L. Toth,
Bernhard Schmidt, Bryan L. Westerman, Hugh T. Morgan, S. S. (2004). Techniques
and Applications of Automatic Tube Current Modulation for CT1 Introduction. RSNA.
Martini, K., Higashigaito, K., Barth, B. K.,
Baumueller, S., Alkadhi, H., & Frauenfelder, T. (2015). Ultralow-dose CT
with tin filtration for detection of solid and sub solid pulmonary nodules: A
phantom study. British Journal of Radiology, 88(1056).
https://doi.org/10.1259/bjr.20150389
McKenney, S. E., Seibert, J. A., Lamba, R., &
Boone, J. M. (2014). Methods for CT automatic exposure control protocol
translation between scanner platforms. Journal of the American College of
Radiology, 11(3), 285–291.
https://doi.org/10.1016/j.jacr.2013.10.014
McKenzie, K. A., & Mahnken, J. D. (2023).
Simulating and estimating agreement in the presence of multiple raters and
covariates. Statistics in Medicine, 42(11), 1687–1698.
https://doi.org/10.1002/sim.9694
Mcnitt-gray, M. (2011). Tube Current Modulation
Approaches : Overview , Practical Issues and Potential Pitfalls Tube
Current Modulation - Overview. Current.
Mozaffary, A., Trabzonlu, T. A., Kim, D., &
Yaghmai, V. (2019). Comparison of tin filter–based spectral shaping CT and
low-dose protocol for detection of urinary calculi. American Journal of
Roentgenology, 212(4), 808–814. https://doi.org/10.2214/AJR.18.20154
Noviliawati, R., Anam, C., Sutanto, H., Dougherty, G.,
& Mak’ruf, M. R. (2021). Automatic validation of the gantry tilt in a
computed tomography scanner using a head polymethyl methacrylate phantom. Polish
Journal of Medical Physics and Engineering, 27(1), 57–62.
https://doi.org/10.2478/pjmpe-2021-0008
Rajendran, K., Voss, B. A., Zhou, W., Tao, S., Delone,
D. R., Lane, J. I., Weaver, J. M., Carlson, M. L., Fletcher, J. G., McCollough,
C. H., & Leng, S. (2020). Dose Reduction for Sinus and Temporal Bone
Imaging Using Photon-Counting Detector CT with an Additional Tin Filter. Investigative Radiology, 55(2), 91–100.
https://doi.org/10.1097/RLI.0000000000000614
Rodrigues, H., Ramos, R.,
Fagundes, L., Galego, O., Navega, D., Coelho, J. d. O., Alves, F. C., &
Cunha, E. (2020). Mastoid, middle ear and inner ear analysis in CT scan
– a possible contribution for the identification of remains. Medicine,
Science and the Law, 60(2), 102–111.
https://doi.org/10.1177/0025802419893424
Romans, L. E. (2011). Computed Tomography for
Technologist. Wolters Kluwer Health.
Schüle, S., Strobel, J. R. B., Lorenz, K. J., Beer,
M., & Hackenbroch, C. (2023). Tin filter compared to low kV protocols -
optimizing sinonasal imaging in computed tomography. PLoS ONE, 18(1
January), 1–13. https://doi.org/10.1371/journal.pone.0279907
SEERAM, E. (2016). COMPUTED TOMOGRAPHY Physical
Principles, Clinical Applications, and Quality Control (FOURTH). Elsevier.
Serhal, C. B., Jacobs, R., Gijbels, F., Bosmans, H.,
Hermans, R., Quirynen, M., & Van Steenberghe, D. (2001). Absorbed doses
from spiral CT and conventional spiral tomography: A phantom vs. Cadaver study.
Clinical Oral Implants Research, 12(5), 473–478.
https://doi.org/10.1034/j.1600-0501.2001.120507.x
Setiawan, A. M. B., Anam, C., Hidayanto, E., Sutanto,
H., Naufal, A., & Dougherty, G. (2023). Comparison of noise-power spectrum
and modulation-transfer function for CT images reconstructed with iterative and
deep learning image reconstructions: An initial experience study. Polish
Journal of Medical Physics and Engineering, 29(2), 104–112.
https://doi.org/10.2478/pjmpe-2023-0012
Shrimpton, P. C., Wall, B. F., & Fisher, E. S.
(1981). The tissue-equivalence of the Alderson Rando anthropomorphic phantom
for X-rays of diagnostic qualities. Physics in Medicine and Biology, 26(1),
133–139. https://doi.org/10.1088/0031-9155/26/1/013
Singh, S., Kalra, M. K., Gilman, M. D., Hsieh, J.,
Pien, H. H., Digumarthy, S. R., & Shepard, J.-A. O. (n.d.). Adaptive
Statistical Iterative Reconstruction Technique for Radiation Dose Reduction in
Chest CT: A Pilot Study 1. Radiology, 259.
https://doi.org/10.1148/radiol.11101450/-/DC1
Sowby, F. D. (1981). ICRP Statement on Tissue
Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs
– Threshold Doses for Tissue Reactions in a Radiation Protection Context. Annals
of the ICRP, 6(1), 1. https://doi.org/10.1016/0146-6453(81)90127-5
Tacelli, N., Remy-Jardin, M., Flohr, T., Faivre, J.
B., Delannoy, V., Duhamel, A., & Remy, J. (2010). Dual-source chest CT
angiography with high temporal resolution and high pitch modes: Evaluation of
image quality in 140 patients. European Radiology, 20(5),
1188–1196. https://doi.org/10.1007/s00330-009-1638-5
Tack, D., Kalra, M. K., & Gevenois, P. A. (2012). Radiation
Dose from Multidetector CT. https://doi.org/10.1007/987-3-642-24535-0
Widyanti, E. R., Anam, C., Hidayanto, E., Naufal, A.,
& Haekal, M. (2023). An evaluation of automated measurement of slice
sensitivity profile of computed tomography image: Field of view variations. Indonesian
Journal of Electrical Engineering and Computer Science, 29(3),
1430–1437. https://doi.org/10.11591/ijeecs.v29.i3.pp1430-1437
Woods, M., & Brehm, M. (n.d.). Shaping the beam
Versatile filtration for unique diagnostic potential within Siemens
Healthineers CT.
Zhang, G. M. Y., Shi, B., Sun, H., Xue, H. D., Wang,
Y., Liang, J. X., Xu, K., Wang, M., Wang, M., Xu, M., & Jin, Z. Y. (2017).
High-pitch low-dose abdominopelvic CT with tin-filtration technique for
detecting urinary stones. Abdominal Radiology, 42(8), 2127–2134.
https://doi.org/10.1007/s00261-017-1103-x
|
©
2023 by the authors. It was submitted for possible open-access publication
under the terms and conditions of the Creative Commons Attribution (CC BY SA) license (https://creativecommons.org/licenses/by -sa / 4 .0/ ). |