IMDOSE PHANTOM DEVELOPMENT FOR QUALITY ASSURANCE AND
CONTROL IN COMPUTED TOMOGRAPHY IMAGING STUDY OF IMAGE QUALITY AND DOSAGE
ASSESSMENT
Djuli Pontjowijono1, Donny Kristanto
Mulyantoro2, Rini Indrati3
Politeknik Kesehatan Kemenkes Semarang,
Central Java, Indonesia
djuli70@gmail.com
KEYWORDS |
ABSTRACT |
Phantom,
CT Scan; image quality; and dose |
C.T. has made
a great contribution to detecting and diagnosing patients' diseases. C.T.,
before being applied to patients, must pass quality assurance and control with
phantom. However,
the availability of phantoms is still minimal because of their high price. His study aims to
create and evaluate cylindrical phantoms for quality assurance and control
procedures based on image quality and dose parameters. The research method
involves creating an image-quality phantom design that references the CatPhan phantom using Sketchup.
The results show that the phantom image provides results from the standard
phantom; the SNR, CNR, MTF, noise, and MTF values of the ImDose
phantom are still within the standard phantom criteria. Based on statistical
tests, the image quality of SNR, CNR, MTF, NPS, and noise is not
statistically significantly different from the standard phantom (p>0.05).
The uniformity value still needs to be improved because it has a significant
difference from the standard phantom with an ap
value <0.05. Increasing the voltage and current gave results proportional
to improving image quality and dose. ImDose
phantoms for quality assurance and C.T. control have been successfully
developed. |
DOI: 10.58860/ijsh.v2i10.105 |
|
Corresponding Author: Djuli Pontjowijono
Email: djuli70@gmail.com
INTRODUCTION
X-rays, such
as in medical imaging, have been widely used in the health sector. X-rays can
enter the body and interact with tissue to produce images of the body's
internal anatomy based on differences in X-ray attenuation coefficients (Ance,
2021) (Agency
& Dance, 2014). One medical imaging
modality that has been widely used is Computed Tomography (C.T.) Scanner. C.T.
Scan is a computerized tomography imaging technique with many-body X-ray
measurements for disease diagnosis and radiotherapy treatment planning (Anam
et al., 2019; Davis et al., 2017; Jeong & Lee, 2014). C.T. scans for diagnosis have been widely used and have
increased sharply in medical imaging. According to the International
Commissioning on Radiological Protection (ICRP), it has been reported that the
frequency of use of C.T. scans is around 10% of all radiological examinations (Task
Group on Control of Radiation Dose in Computed Tomography, 2000) and will continue to increase from year to year. C.T. scan image
reconstruction is based on the attenuation difference value of the interaction
of X-rays with tissue in the form of charge signal data known as the C.T.
number on the Hounsfield Unit (H.U.) scale (Anam
et al., 2016). This H.U. value is used
to assess the disease suffered by the patient from the scanned tissue. This
H.U. can be influenced by several factors, such as scanner design and
calibration, reconstruction algorithm, beam hardening artefacts, body size,
object orientation, and tube voltage (Hunter
et al., 1983; Levi et al., 1982; Samei & Pelc, 2019). These factors cause C.T. number variability, which can
significantly affect the reliability of the data produced by the C.T. scan.
Apart from the image quality aspect, you also need to pay attention to the dose
received by the body. The dose received by the body must comply with the ALARA
principle; namely, the dose received must be as minimal as possible without
obscuring the resulting image information. Therefore, CT Scans must be managed
through quality assurance and quality control (Q.C.).
Radiation
protection authorities periodically perform C.T. scan acceptance tests and QA /
QC tests to assess radiation dose and image quality. QA / QC testing of image
quality and radiation dose is mandatory. QA / QC procedures related to image
quality and dose in the C.T. Scan modality consist of quantitative assessment
with a standard phantom. Quality Assurance (Q.A.) includes all procedures to
ensure equipment meets quality requirements. In diagnostic radiology, Q.A. aims
to perform the most appropriate X-ray examination for diagnosis with optimized
exposure factors, providing high-quality images at acceptable doses.
Meanwhile, QC
involves a process through which relevant performance parameters are measured
and compared with existing standards, basic values, and accepted tolerances (Agency
& Dance, 2014). The most important
parameters for C.T. image quality are image noise, CT-number uniformity,
spatial resolution, and low contrast resolution. Meanwhile, CT dose parameters
use the C.T. Dose Index (CTDI) value [10,
11]. The quality assurance
process certainly cannot use humans as test objects; of course, this is a loss
for humans. Therefore, QA / QC procedures require a replacement object called a
phantom.
Fantom is a real imitation model that has parity with the real network. The equality
referred to is that the attenuation value of the phantom material tissue is
equivalent to the real tissue (White,
1993). Standard phantoms for QA
/ QC are generally made from polymethyl methacrylate
(PMMA) material. This standard PMMA phantom has two diameters, one with a
diameter of 16 cm representing the head and another with a diameter of 32 cm to
replace the patient's body. This phantom is cylindrical with a length of 15 cm.
This phantom has five holes in the centre and four others at the edges (Mubarok
et al., 2016). However, this standard
phantom is very expensive, and many hospitals in Indonesia still need it. Fantom is very useful for independent quality assurance,
training and education. Several studies related to C.T. phantom materials have
been carried out (Sookpeng
et al., 2016) used nylon as head and
body phantom material to measure CTDI and reported that the C.T. number value
was around 20 H.U. lower than standard PMMA phantoms (Sookpeng
et al., 2016). Apart from that (Hilmawati
et al., 2020) made a CTDI phantom from
methyl ethyl ketone peroxide (MEKP), which gave an H.U. value of 1 – 9% greater
than the PMMA phantom. The two phantoms are still for testing the quality of
radiation doses, there is not yet a phantom package to guarantee image quality
and dose. Apart from that, the level of precision during manufacture still
allows for errors. Therefore, this research will propose the ImDose Image and Dose phantom to guarantee image quality
and radiation dose using Polylactic Acid (PLA)
material using 3D printing technology. PLA material is similar to human tissue,
with a density of around 1 – 1.23 g/cm3 [15].
This research
has several benefits to be achieved, namely, Providing
research references in the development aspect of the ImDose
phantom for quality assurance and control in X-ray modalities. This will help
increase understanding of ImDose phantom development
and the importance of quality assurance and control.
Provide
knowledge about suitable materials for phantoms, quality assurance processes,
and controls that can be carried out independently. With this knowledge, this
research will improve the ability to select appropriate materials for phantoms
and effectively carry out quality assurance and control processes.
They provide
independence for affordable medical equipment in Indonesia through education,
training, quality assurance and control processes for hospital imaging
equipment. This will help increase access to education and training in medical
imaging and help hospitals in Indonesia maintain the quality and control of
their imaging equipment at an affordable cost.
The objectives
of this research are divided into general objectives and specific objectives
covering various aspects. This research aims to create a prototype ImDose phantom that can be used for quality assurance and
control in radiological examinations, with assessment based on image quality
and dose in the same examination protocol. This research will also examine the
effect of voltage and current on the image quality and dose produced by the ImDose phantom.
The specific
objectives of this research include:
a) Design an ImDose phantom design that meets quality assurance and
control standards for image and dose quality assessment.
b) I am realizing the phantom
design into a prototype using 3-dimensional printing technology.
c) Evaluate the image quality
of the ImDose phantom using the parameters Hounsfield
Unit (H.U.), Signal Noise to Ratio (SNR), noise level (noise), contrast to
noise level (CNR), Modulation Transfer Function (MTF), and Noise Power Spectrum
(NPS) at a tube voltage of 120 kV and a current of 200 mA, and compare it with
a standard phantom.
d) Investigating the effect
of variations in tube voltage (90 kV, 120 kV, and 140 kV) on the image quality
of the ImDose phantom using the parameters Signal
Noise to Ratio (SNR), noise level (noise), Contrast to Noise Level (CNR),
Modulation Transfer Function (MTF), and Noise Power Spectrum (NPS).
e) Evaluate the impact of
changes in tube current (50 mA, 200 mA, 300 mA) on the
image quality of the ImDose phantom using the
parameters Signal Noise to Ratio (SNR), noise level (noise), Contrast to Noise
Level (CNR), Modulation Transfer Function (MTF), and Noise Power Spectrum (NPS)
at a tube voltage of 120 kV.
f)
I compared the CT Dose and CT Dose Volume Index (CTDIvol) produced by the ImDose
phantom with a standard phantom at a voltage of 120 kV and a current of 200 mA.
g) Assess the CT Dose and CT
Dose Volume Index (CTDIvol) of the ImDose phantom with variations in tube voltage (90 kV, 120
kV, and 140 kV) and tube current variations (50 mA, 200 mA, 300 mA) at a
voltage of 120 kV.
METHOD
This research is a research and development type research by
making a CT-Scan phantom prototype for quality assurance and control of image
quality and dose. The samples for this study were six image samples and dose
data from the ImDose phantom and standard phantom obtained from the C.T. Scan
modality at the Radiology Installation at Simpang
Lima Gumul Hospital, Kediri. The data was then analyzed for image quality by performing ROIs of the same
size ten times on different slices and
doses.
RESULTS AND DISCUSSION
ImDose phantoms are made using 3D Printing with the filament deposition method.
3D Printing uses a 0.4 mm nozzle with an accuracy of 0.2 mm. The filament is
made using Polylactic Acid material with a filling
percentage of 100% and a wall thickness of 2 mm. The results of creating the ImDose phantom and topogram image
are shown in Figure 1.
(a) |
(b) |
(c) |
Figure 1Results of (a) image and dose phantom fabrication
using 3D Printing, (b) preparation of the ImDose
phantom, (c) C.T. tomogram image of the ImDose
phantom.
The results of
making the ImDose phantom have been successful using
a 3D Printed machine. In addition, the ImDose phantom
can be read using a C.T. scan. For the next discussion, we will compare the
image quality of the ImDose phantom with the standard
phantom from Philips.
Imdose Phantom Image Results and
Standard Phantoms
The standard phantom
image results were taken using a Philips Healthcare MX 16 Slice CT Scanner
using a voltage of 120 kV, 200 mA, as shown in Figure 1(b). Meanwhile, for ImDose phantom C.T. data collection, there are five phantom
image data with C.T. settings, namely 90, 120, and 140 kV at a current of 50 mA
(3 sample groups), 200 and 300 mA at a voltage of 120 kV (2 sample groups). In
the first part, standard phantom and ImDose image
data were compared at the same voltage, 120 kV, 200 mA, and evaluated in 5 slices (Anam
et al., 2022)
(a) (b)
Figure 2ImDose 16 cm phantom
image results and standard phantom
From
Figure 1, qualitatively phantom (a) has image results that are similar to
standard phantoms. The salient difference between Figures 1(a) and (b) is the
skin sheath of the phantom. The ImDose phantom lacks
a high-density outer layer, while the standard phantom uses a high-density
outer layer. Meanwhile, the results of the Fantom
Image with variations in voltage and current are shown in Figure 3.
(a) |
(b) |
(c) |
|
(e) |
Figure 3ImDose phantom images
(a) voltage 90 kV, 50 mA; (b) voltage 120 kV, 50 mA; (c) voltage 140 kV, 50 mA;
(d) voltage 120 kV, 200 mA; and (e) voltage 120 kV, 300 Ma
H.U. value between Imdose phantom and Standard phantom
The
H.U. value of the Imdose phantom with the standard
phantom is taken using a voltage of 120 kV, 200 mA. Calculation of H.U. phantom
values using Radiant Viewer software. This is part of the QA/QC testing
regarding the H.U. value of low to high-density materials. H.U. measurements
were carried out by making ROI on ten different slices. The ROI area for each
slice is made the same. The calculation results are shown in Table 1
Table 1Insertion number values of the Imdose
phantom and standard phantom
Phantom Type |
Insert |
H.U. value |
|
Average |
Range |
||
ImDose Fantom |
Gypsum |
462.74 |
439 to 889 |
PLA 90% |
-65.89 |
-150 to -52 |
|
PLA 75% |
-108.96 |
-203 to -95 |
|
PLA 50% |
-379.38 |
-474 to -388 |
|
PLA 25% |
-509.49 |
-710.7 to -689 |
|
PLA 15% |
-651.27 |
-848 to -768 |
|
PMMA |
25.08 |
0 to 40 |
|
Agarose |
12.6 |
7 to 17 |
|
Air |
-937.12 |
-944 to -922 |
|
PLA100% |
43, 35 |
-12 to 109 |
|
Philips Standard Fantom |
Teflon |
913.4 |
886 to 936 |
Lexan |
97.22 |
72 to 123 |
|
Perspex |
123.48 |
96 to 143 |
|
Polyethylene |
-77.92 |
-111 to -52 |
|
Water |
0 |
-10 to 10 |
|
Aculon |
88.54 |
71 to 107 |
|
Acrylic/PMMA |
3, 54 |
-75 to 74 |
|
Organs (Kalender et al.,
2024; Mar’Ie et al., 2020) |
Soft tissue |
- |
-199 to 120 |
Bone |
- |
121 to 1600 |
|
Lungs |
- |
-949 to -120 |
|
Water |
- |
-1000 to -950 |
From Table 1, the ImDose
phantom value has an average H.U. value range from 462.74 to -937.12.
Meanwhile, the standard phantom only has an average value ranging from -77.92
to 913.4. The ImDose phantom generally has a larger
range value than the standard phantom. This is a positive thing because the
density of human organs is from -949 (lung organs) to 1600 (dense bones).
The H.U. between the ImDose phantom insert, and the
standard phantom has a difference of 15.43% for PLA, 90% with Polyethene, and 7% for agarose
with water. For the range of H.U. values between ImDose
phantoms, Gypsum represents Teflon, PMMA represents lexan,
Perspex, and aculon, and agarose
represents water. Standard phantoms have higher-density inserts than ImDose phantoms. In general, the Imdose
phantom can be used to test H.U. values.
Uniformity value of Imdose Fantom H.U. and Standard Fantom
The
H.U. uniformity value between ImDose phantoms is
calculated by determining the H.U. value at 4 points at the edge and 1 point in
the middle, as shown in Table 2. Apart from that, testing for uniformity of
H.U. values was also carried out for several slices at the midpoint. The CT
uniformity value between the Imdose phantom and the
standard phantom was tested using a voltage of 120 kV, 200 mA. The results of
calculating the uniformity value are shown in Figure 4.
The centre and average H.U. of
the four edges of each phantom are displayed. The edge values presented are the
average values of the four edge points' four positions.
Uniformity is the standard
deviation of the centre and edge areas.
Fantom |
Slice |
Nilai HU |
Standar deviasi |
|
tengah |
Rata-rata area tepi |
|||
Fantom ImDose |
1 |
69,27 |
62,68 |
4,66 |
2 |
55,65 |
52,04 |
2,55 |
|
3 |
58,56 |
55,04 |
2,49 |
|
4 |
58,18 |
56,98 |
0,85 |
|
5 |
49,86 |
46,79 |
2,17 |
|
Fantom standar |
1 |
3,623 |
1,86 |
1,25 |
2 |
3,654 |
1,92 |
1,23 |
|
3 |
2,985 |
1,96 |
0,72 |
|
4 |
3,412 |
1,94 |
1,04 |
|
5 |
2,698 |
2,04 |
0,46 |
Figure 4Imdose vs standard phantom uniformity
Table 2 shows the uniformity values
of ImDose and standard phantoms. Fantom
ImDose has a uniformity in the range of 0.85 – 4.66.
Meanwhile, standard phantoms have uniformity in the range of 0.46 – 1.25. The
uniformity value is best if the standard deviation value is smaller.
From Figure 4,
the Imdose phantom has good uniformity values with a
constant curve. This means that the H.U. value between slices does not have a significant difference. Meanwhile, the H.U. value between
the ImDose phantom compared
to the standard phantom is much higher and has a difference of 56.72. However,
the H.U. value is between -119 to 120, including
soft tissue (Kalender et al., 2024).
This means that the H.U. value between the two phantoms still reflects the soft
tissue value.
ImDose phantom Signal Noise to
Ratio (SNR) value
The SNR value calculation is obtained from the ratio
of the average H.U. value to the average standard deviation. The ROI was
selected according to the phantom size and was carried out ten times. The
results of the average SNR value between Imdose
phantom vs standard phantom and variations in voltage
and current are shown in Table 3.
Table 3SNR calculation results on the ImDose
phantom with variations in voltage and current
Phantom |
Voltage (kV) |
Current (mAs) |
SNR |
ImDose Fantom |
90 |
50 |
8.25 |
120 |
50 |
17.5 |
|
140 |
50 |
23.3 |
|
120 |
200 |
22.1 |
|
120 |
300 |
25.4 |
|
Standard Fantom |
120 |
200 |
21.7 |
From Table 3 it can be seen that the highest SNR value
is 25.4 at a voltage of 120 kV and a current of 300 mA, while the lowest SNR
value is 8.25 at a voltage of 90 kV and 50 mA. Meanwhile, the SNR value of the ImDose phantom and standard phantom has a difference of
1.84%.
Contrast
Noise to Ratio (CNR) Values of ImDose Fantom and Standard Fantom
The Contrast Noise to Ratio
(CNR) values of the ImDose phantom
and Standard phantom were carried out at a voltage of 120 kV and a current of
200 mA. The results of the CNR calculation are shown in Table 4. Contrast to Noise Ratio (CNR)
measures how contrast (well) the target/lesion can be distinguished from the
background. The CNR value is determined by selecting an ROI between the inset
target area placed in the centre and the surrounding area/ background (top, right, left) for each target. The targets chosen for the inserts
are gypsum and Teflon targets.
Table 4CNR Value Calculation Results
Phantom |
Target |
Voltage (kV) |
Current (mAs) |
CNR |
ImDose Fantom |
|
90 |
50 |
19.58 |
|
120 |
50 |
20.65 |
|
Gypsum |
140 |
50 |
24.60 |
|
|
120 |
200 |
25.24 |
|
|
120 |
300 |
27,26 |
|
PLA90% |
90 |
50 |
-2, 33 |
|
120 |
50 |
-2, 53 |
||
140 |
50 |
-3, 73 |
||
120 |
200 |
-5, 83 |
||
120 |
300 |
-6, 46 |
||
PMMA |
90 |
50 |
4, 49 |
|
120 |
50 |
6, 29 |
||
140 |
50 |
8, 33 |
||
120 |
200 |
9, 44 |
||
120 |
300 |
10 , 16 |
||
Agarose |
90 |
50 |
3 , 86 |
|
120 |
50 |
7, 23 |
||
140 |
50 |
9, 19 |
||
120 |
200 |
8, 33 |
||
120 |
300 |
9, 48 |
||
PLA 75% |
90 |
50 |
-5, 56 |
|
120 |
50 |
-6, 41 |
||
140 |
50 |
-7, 92 |
||
120 |
200 |
-7, 39 |
||
120 |
300 |
-8 , 12 |
||
Standard Fantom |
Teflon |
120 |
200 |
47.63 |
Lexan |
120 |
200 |
15, 06 |
|
Perspex |
120 |
200 |
17, 84 |
|
Polyethylene |
120 |
200 |
-6, 02 |
Based on Table 4, it is known that in the ImDose phantom, the gypsum insert has a CNR value in the
range of 19.58 to 27.26, the 90% PLA insert has a CNR value from -2.33 to
-6.46, the PMMA insert has a CNR value of 4, 49 to 10.16, the agarose insert has a CNR of 3.86 to 9.48, and 75% PLA has a
value range of -5.56 to -8.12. When compared with the inserts in standard
phantoms, gypsum inserts have a difference of 22.39 with Teflon inserts, PLA
90% inserts have a difference of 0.22 with Polyethene,
PMMA inserts have a difference of 8.4 with Perspex, and agarose
inserts have a difference of 8.4 with Perspex. 6.73 of Lexan inserts
at 120 kV voltage, 200 mA current.
Fantom Imdose
and Standard Fantom Modulation Transfer Function (MTF) values
Analysis of
MTF results from ImDose phantom images using IndoQCT software. The results of the MTF analysis are shown
in Figure 5.
(a) |
(b) |
Figure 5MTF calculation results between the ImDose Fantom and
the standard Philips Fantom
Based on
testing, the spatial resolution MTF value of the ImDOSE
and standard phantoms is equivalent at MTF50%, namely 0.06 and MTF 10%, which
has a difference of 0.11 greater than the ImDose
phantom. The greater the MTF value, the greater the phantom has higher spatial
resolution and good image results. Here, it was found that the ImDose phantom had better resolution than the standard
Philips phantom. The results of the ImDose phantom
evaluation based on voltage and current are shown in Table 5.
Table 5MTF Value Calculation Results
Phantom |
Voltage (kV) |
Current (mA) |
MTF |
|
50% |
10% |
|||
ImDose Fantom |
90 |
50 |
0.06 |
0.17 |
120 |
50 |
0.08 |
0.22 |
|
140 |
50 |
0.15 |
0.33 |
|
120 |
200 |
0.17 |
0.24 |
|
120 |
300 |
0.2 |
0.37 |
|
Standard Fantom |
120 |
200 |
0.06 |
0.11 |
Fantom Imdose
and Fantom Standard Noise Power Spectrum Values
The results of the NPS value analysis in the first part will compare the
NPS value between the ImDose phantom and the standard
shown in Figure 6.
(a) |
(b) |
Figure 6Evaluation results of NPS values between (a) ImDose phantom; (b) standard
phantom
Table 6NPS Value Calculation Results
Phantom |
Voltage (kV) |
Current (mA) |
NPS |
|
Noise (H.U.) |
NPS Integral |
|||
ImDose Fantom |
90 |
50 |
15.05 |
16.55 |
120 |
50 |
13.22 |
15.63 |
|
140 |
50 |
10.52 |
12.52 |
|
120 |
200 |
7.64 |
9.23 |
|
120 |
300 |
6.9 |
8.35 |
|
Standard Fantom |
120 |
200 |
8.04 |
10.98 |
Apart from
that, from Table 6, the noise value decreases as the voltage and current
increase. This is also by the NPS value, the higher the voltage and current,
the lower the NPS value. This indicates that the image quality improves as the
voltage and current increase.
Thus, the ImDose phantom has been successfully created, with image
quality equivalent to a standard Philips phantom. Apart from
that, the higher the voltage and current, the better the image quality.
However, this must pay attention to the dose received by the patient. Dose
evaluation will be discussed in subchapter 9.
Standard phantom CTDIvol and Imdose values
CTDIvo measurements used a standard phantom head with a
diameter of 16 cm owned by the Surabaya Health Facilities Security Center (HFSC) on September 6 2023. Measurements were
carried out at 4 points at the edges and 1 point in the centre. Head phantom
CTDI measurement results with parameters 120 kV, 200 mA, and slice thickness 5
mm. The results of standard head phantom CTDI measurements are shown in Table 7.
Table 7Measurement and calculation
results CTDI Vol standard
head phantom
Location |
Measurement results |
||
Dosage (mGy) |
CTDIw |
CTDIvol |
|
Center |
33.01 |
34.84 |
51,751 |
Edge 1 |
38.23 |
||
Edge 2 |
35.01 |
||
Edge 3 |
36.45 |
||
Edge 4 |
33.33 |
Meanwhile,
to measure C.T. dose and CTDIvol from the ImDose phantom using the same scanning parameters as the ImDose phantom. The results of measuring and
calculating C.T. dose and CTDIvol from the ImDose phantom at a voltage of 120 kV and 200 mA are shown
in Table 8.
Table 8CTDIvol measurement and calculation
results from the ImDose phantom
Location |
Measurement results |
||
Dosage (mGy) |
CTDIw |
CTDIvol |
|
Center |
29.45 |
30.32 |
45,259 |
Edge 1 |
30.66 |
||
Edge 2 |
30.78 |
||
Edge 3 |
32.57 |
||
Edge 4 |
29.03 |
From Tables 7
and 8, the C.T. dose from 5 measurement points shows that the C.T. dose from
the standard phantom is greater than the ImDose
phantom. The average difference in dose measurements at 5 points was 12.88%.
This difference can be caused by several factors, namely a placement error, and
perhaps the height of the C.T. table also has an influence. Furthermore, based
on the CTDIvol value, the standard phantom and ImDose phantom have values of 51.751 and 45.259. From these
results, the difference between standard phantom and ImDose
is 14.34%. This difference is still within the tolerance range. The threshold
for CTDI head phantom suitability testing is at most 20%. These results
indicate that the ImDose phantom has performance or
capabilities that can be used for C.T. dose index measurements. Furthermore, CT
dose and CTDIvol measurements were carried out for
various voltage and current variations, as shown in Tables 9 – 12
CT dose and CTDIvol phantom ImDose values for
voltage and current variations
The results of C.T. dose and CTDIvol phantom ImDose measurements at various voltages and currents are
shown in Table 9 – 12. The results show that the greater the voltage
simultaneously, the higher the C.T. dose. This also applies to C.T. doses with
current variations at the same voltage.
Table 9CTDI Vol phantom
ImDose Measurement Results
for voltage 90 kV and 50 mA
Location |
Measurement results |
||
Dosage (mGy) |
CTDIw |
CTDIvol |
|
Center |
2, 960 |
3, 23 |
4, 196 |
Edge 1 |
3,666 |
||
Edge 2 |
3,389 |
||
Edge 3 |
3,416 |
||
Edge 4 |
2,996 |
Table 10 CTDI Vol phantom
ImDose Measurement Results
voltage 120 kV and 50 mA
Location |
Measurement results |
||
Dosage (mGy) |
CTDIw |
CTDIvol |
|
Center |
6,891 |
7, 28 |
9, 455 |
Edge 1 |
8,013 |
||
Edge 2 |
7,523 |
||
Edge 3 |
7,584 |
||
Edge 4 |
6,782 |
Table 11 ImDose phantom CTDIVol measurement
results of 140 kV and 50 mA
Location |
Measurement results |
||
Dosage (mGy) |
CTDIw |
CTDIvol |
|
Center |
9,663 |
10, 37 |
13, 470 |
Edge 1 |
11.38 |
||
Edge 2 |
10.95 |
||
Edge 3 |
10.77 |
||
Edge 4 |
9,805 |
Table 12ImDose phantom CTDIVol measurement results
for voltage 120 kV and 300 mA.
Location |
Measurement results |
||
Dosage (mGy) |
CTDIw |
CTDIvol |
|
Center |
40.4 |
56, 53 |
54, 00 |
Edge 1 |
48.11 |
||
Edge 2 |
44.64 |
||
Edge 3 |
45.89 |
||
Edge 4 |
41.73 |
The results show that the
measurement above has the largest dose; the right, left, and centre edges show
close dose results. Meanwhile, the deepest point provides the smallest dose.
So, the deeper the dose decreases due to the thickness factor. The comparison
of the CTDIvol dose measured with the ImDose phantom with the indexed Console results is shown in
Table 13.
Table 13CTDIVol comparison results of ImDose
phantom measurements with console
Tool settings |
CTDIvol (mGy) |
Difference (%) |
||
kV |
mA |
Console |
Measurement |
|
90 |
50 |
3, 720 |
4, 196 |
12.80 |
120 |
50 |
8,790 |
9, 455 |
7.57 |
140 |
50 |
13,010 |
13, 470 |
3.54 |
120 |
300 |
51.45 |
56,530 |
9.87 |
Table
13 shows the CTDIvol comparison between measurements
with the ImDose phantom and console. Table 4.13 shows
that the largest difference in CTDIvol is 12.80%, and
the smallest is 3.54%. These results show that the ImDose
phantom can work with variations in voltage and current.
Statistical Test Analysis
of ImDose Fantom with
Standard Fantom
The ImDose HU phantom and standard phantom statistical tests
used the independent T-test statistical test method with a significance value
0.05. Tests were carried out on data between the ImDose
phantom and a standard phantom with a voltage of 120 kV, 200 mA. Before the
data is tested, the independent t-test must look at the normality distribution
of the data. The data normality test uses Shapiro Wilk
because this research data is below 30. The data normality test uses a
significance value of 0.05. The test results are shown in Table 1 four below.
The ImDose fathom normality distribution results for
all parameters were normally distributed with a p-value> 0.05. Meanwhile,
the standard phantom normality test results were normally distributed for all
parameters with a p-value> 0.05. Except for the standard phantom SNR
parameters, it is not normally distributed with a p-value < 0.05.
Table 14normality tests data from image quality parameters
on the
ImDose phantom and Standard phantom at voltages of
120 kV and 200 mA
Image quality test parameters |
ImDose Fantom |
Standard Fantom |
Sig. |
Sig |
|
Uniformity |
0.122 |
0.494 |
SNR |
0.230 |
0.017 |
CNR |
0.575 |
0.851 |
MTF |
0.389 |
0.057 |
NPS |
0.982 |
0.079 |
Noise |
0.864 |
0.945 |
CTDI Dosage |
|
0, 753 |
Next, an independent
t-test was carried out because the test was conducted to ensure the normality
assumption was met. The results of the statistical independent t-test analysis
are shown in Table 15.
Table 15t tests for image quality
between the ImDose phantom and the
Standard phantom at 120 kV and 200 mA.
Image
quality and dose test parameters |
p-value |
Uniformity |
0.002 |
SNR |
0.988 |
CNR |
0.291 |
MTF |
0.294 |
NPS |
0.654 |
Noise |
0.053 |
CTDI Dose |
0.585 _ |
CONCLUSION
Based on the
results of the analysis that has been carried out, several things can be
concluded as follows: The ImDose phantom was
successfully created using 3D printing technology and can be used to measure
image quality and dose in one phantom. ImDose Fantom has a Hounsfield Unit (H.U.) value in the range of
soft tissue H.U. values. The results of the intercomparison
of the H.U. value of the ImDose phantom with the
standard phantom show that the H.U. value of the ImDose
phantom is higher than the standard phantom. This result is also statistically
significant, with a p-value < 0.05, indicating a significant difference in
the uniformity of H.U. values between the ImDose
phantom and the standard phantom. Image quality parameters such as
Signal-to-Noise Ratio (SNR), Contrast-to-Noise Ratio (CNR), Modulation Transfer
Function (MTF), noise level (noise), and Noise Power Spectrum (NPS) on the ImDose phantom show no difference, which is statistically
significant (p-value > 0.05). However, the uniformity parameter has a
significant difference, with a p-value <0.05.
ImDose phantom quality parameters, such as SNR, CNR, and MTF, further improve
with increasing current and voltage. This indicates that the phantom can
produce better images with a clearer background. In addition, the noise level
and NPS tend to decrease as the voltage and current increase, indicating an
increase in image quality. The C.T. dose and CTDIvol
testing results between the ImDose phantom and the
standard phantom showed a difference of 12.88% for the C.T. dose and 14.34% for
the CTDIvol dose. The CT dose in the standard phantom
is higher than in the ImDose phantom. The effect of
voltage and current on C.T. dose and CTDIvol shows an
increase in dose values with increasing voltage and current. However, the
difference between measurements and CTDIvol was still
below 20%, indicating good consistency in dose measurements. These results
provide a clear picture of the ImDose phantom's
ability to measure image quality and dose and the impact of voltage and current
variables on measurement results.
REFERENCES
agency,
I. A. E., & Dance, D. R. (2014). Diagnostic
Radiology Physics: A Handbook For Teachers And Students. International
Atomic Energy Agency.
Anam,
C., Budi, W. S., Adi, K., Sutanto, H., Haryanto, F., Ali, M. H., Fujibuchi, T.,
& Dougherty, G. (2019). Assessment Of Patient Dose And Noise Level Of
Clinical Ct Images: Automated
Measurements. Journal Of
Radiological Protection : Official Journal Of The Society For Radiological Protection, 39(3), 783–793.
Https://Doi.Org/10.1088/1361-6498/Ab23cc
Anam,
C., Haryanto, F., Widita, R., Arif, I., & Dougherty, G. (2016). Automated
Calculation Of Water-Equivalent Diameter (Dw) Based On Aapm Task Group 220. Journal
Of Applied Clinical Medical Physics, 17(4), 320–333. Https://Doi.Org/10.1120/Jacmp.V17i4.6171
Anam,
C., Naufal, A., Fujibuchi, T., Matsubara, K., & Dougherty, G. (2022).
Automated Development Of The Contrast–Detail Curve Based On Statistical
Low‐Contrast Detectability In Ct Images. Journal Of Applied Clinical Medical Physics, 23(9), E13719.
Ance,
R. F. (2021). Prosedur Pemeriksaan
Ct-Scan Abdomen Kontras Pada Klinis Kanker Serviks Di Instalasi Radiologi Rsud
Arifin Achmad Provinsi Riau.
Davis,
A. T., Palmer, A. L., & Nisbet, A. (2017). Can Ct Scan Protocols Used For
Radiotherapy Treatment Planning Be Adjusted To
Optimize Image Quality And Patient Dose? A Systematic Review. The British Journal Of Radiology, 90(1076), 20160406.
Https://Doi.Org/10.1259/Bjr.20160406
Gulliksrud,
K., Stokke, C., & Trægde Martinsen, A. C. (2014). How To Measure Ct Image
Quality: Variations In Ct-Numbers, Uniformity And Low Contrast Resolution For A
Ct Quality Assurance Phantom. Physica
Medica, 30(4), 521–526.
Https://Doi.Org/10.1016/J.Ejmp.2014.01.006
Hilmawati,
R., Sutanto, H., Anam, C., Arifin, Z., Asiah, R. H., & Soedarsono, J. W.
(2020). Development Of A Head Ct Dose Index (Ctdi) Phantom Based On Polyester
Resin And Methyl Ethyl Ketone Peroxide (Mekp): A Preliminary Study. Journal Of Radiological Protection, 40(2), 544.
Https://Doi.Org/10.1088/1361-6498/Ab81a6
Hunter,
T. B., Pond, G. D., & Medina, O. (1983). Dependence Of Substance Ct Number
On Scanning Technique And Position Within
Scanner. Computerized
Radiology : Official Journal Of The Computerized Tomography Society,
7(3), 199–203.
Https://Doi.Org/10.1016/0730-4862(83)90099-9
Jeong,
J. E., & Lee, S. J. (2014). Performance Comparison Of Ray-Driven System
Models In Model-Based Iterative Reconstruction For Transmission Computed
Tomography. Journal Of Biomedical
Engineering Research, 35(5),
142–150. Https://Doi.Org/10.9718/Jber.2014.35.5.142
Kalender,
W. A., Schmidt, B., & Flohr, T. (2024). Kalender - Computed Tomography: Fundamentals, System Technology, Image
Quality, Applications. Wiley.
Levi,
C., Gray, J. E., Mccullough, E. C., & Hattery, R. R. (1982). The
Unreliability Of Ct Numbers As Absolute Values. Ajr. American Journal Of Roentgenology, 139(3), 443–447. Https://Doi.Org/10.2214/Ajr.139.3.443
Mar’ie,
K., Lestariningsih, I., Nurlely, & Soejoko, D. S. (2020). Phantom Design
For Analysis Of Ct Image Quality From Single-Source And Dual-Source Ct Scan. Journal Of Physics: Conference Series,
1568(1).
Https://Doi.Org/10.1088/1742-6596/1568/1/012019
Mubarok,
S., Lubis, L. E., & Pawiro, S. A. (2016). Parameter-Based Estimation Of Ct
Dose Index And Image Quality Using An In-House Androidtm-Based
Software. Journal Of Physics:
Conference Series, 694(1),
12037. Https://Doi.Org/10.1088/1742-6596/694/1/012037
Samei,
E., & Pelc, N. J. (2019). Computed
Tomography: Approaches, Applications, And Operations. Springer
International Publishing.
Sookpeng,
S., Cheebsumon, P., Pengpan, T., & Martin, C. (2016). Comparison Of
Computed Tomography Dose Index In Polymethyl Methacrylate And Nylon Dosimetry Phantoms. Journal Of Medical Physics, 41(1), 45–51. Https://Doi.Org/10.4103/0971-6203.177287
Task
Group On Control Of Radiation Dose In Computed Tomography. (2000). Managing
Patient Dose In Computed Tomography. A Report Of The International Commission
On Radiological Protection. Annals Of
The Icrp, 30(4), 7–45.
Https://Doi.Org/10.1016/S0146-6453(01)00049-5
White,
D. R. (1993). The Design And Manufacture Of Anthropomorphic Phantoms. Radiation Protection Dosimetry, 49(1–3), 359–369.
Https://Doi.Org/10.1093/Rpd/49.1-3.359
|
©
2023 by the authors. It was
submitted for possible
open-access publication under
the terms and conditions of the Creative Commons Attribution (CC BY SA) license
(https://creativecommons.org/licenses/by-sa/4.0/). |